WE2107 - Communication commands
70
I1820-1.3 enHBM
Tare Set
(Gross/net selection)
Property Content Note
Command string TAS
No. of parameters 1
Parameter range P1= 0/1 (0=net, 1=gross)
Factory default 1
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
no
Parameter backup With command TDD1
Input Master TAS(P1); No response
Query Master TAS?;
Response WE P1 crlf P1 = 1 character
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The command changes data output over (MSV?
).
TAS0:net measured value
The value in tare memory is subtracted from the current measured value.
TAS1:gross measured value
The value in tare memory is not offset. The tare value is unchanged during the
gross/net changeover.