WE2107 - Communication commands
40
I1820-1.3 enHBM
Manual input of the nominal (rated) value via LWT:
1. Use the SPW command to enter your password.
2. Set legal for trade switch to LFT=0
3. Reset the correction factor: ACA==ACU
4. the scale is unloaded.
5. Query the measurement output (MIV
?)
6. Enter the LDW value.
7. Load the scale with max. capacity.
8. Query the measurement output (MIV
?)
9. Use the command LWT<nominal (rated) value> to enter the measured value for nominal
load.
The value entered is stored and offset with the LDW value previously measured or
entered.
10.Save the new curve with TDD1;.
NOTE
The characteristic curve commands LDW and LWT should be entered or executed in the
following order: LDW followed by LWT. The input data is only offset if these two parameters
have been entered or measured in pairs.
When the factory default curve is measured with LDW/LWT, this resets the user curve to the
default values LDW=0, LWT=200000, CWT
=100000 and ACU:=ACA
With partial load adjustment (measurement):
The LWT value is converted to 100 % in accordance with the CWT value entered and the
CWT value is then reset to its 100 % value (=100000).