WE2107 - Communication commands
I1820-1.3 en HBM
External display end character 1
(Protocol external display definition of the end character 1)
Property Content Note
Command string ED1
No. of parameters 1
Parameter range 0 31
Factory default 13
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
No
Parameter backup With command TDD1
Input Master ED1(P1); No response
Query Master ED1?;
Response WE xx crlf 2 characters
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
This command defines the first end character in the output string.
ED1=0: no end character is defined (the telegram is 1 byte shorter)
ED1=131 an end character is defined ( reference ASCII‐chart PC437)
Examples:
ED1=3: end character is ETX (=03hex)
ED1=13: end character is an CR (=0d hex)