WE2107 - Communication commands
68
I1820-1.3 enHBM
Tare Value
(Tare value)
Property Content Note
Command string TAV
No. of parameters 1
Parameter range P1= 0...±99999
Factory default 0 disabled
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
No
Parameter backup With command TDD1
Input Master TAV(P1); No response
Query Master TAV?;
Response WE X crlf
(X= current 7‐digit tare value with sign)
X = 6 chars.
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The tare value can be pre‐assigned, or a tare value saved by the tare function (TAR
) is
output.
The value is on the LDW/LWT curve (0...NOV) scaled with the NOV
parameter. After making
the curve inputs with the commands SZA, SFA
or LDW, LWT, the tare memory is cleared
(content=0).
When the tare value is entered, the display switch to NET display weight.
When the tare value is input, the net value is identified in the displayas a pretare value (PT).
TAV