127
WE2107 - Communication commands
I1820-1.3 en HBM
Limit values
(Limit values switches)
Property Content Comment
Command string LIV
No. of parameters 1
Parameter range P1= 1…4, P2=0,1,2, P3=0/1
P4=P5= 0...+
99999
Factory setting x, 0, 0, 0, 0 for all limit values deactivated
Response time < 15 ms
Password protection No
Relevant for legal for
trade
No
Parameter backup With command TDD1
Master input LIV(P1),<P2>,<P3>,<P4>,<P5>; No response
Master query LIV(P1)?;
WE response P2,P3,P4,P5 crlf P2,P3= 1 charac
ter,
P4,5 = each
6 characters
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The WE contains four limit value switches with selectable hysteresis. These can monitor
gross or net measured values. The monitoring speed depends on the set output rate. Moni
toring is always implemented, even when there is no communication via the serial interface.
LIV