57
WE2107 - Communication commands
I1820-1.3 en HBM
Zero Tracking
(Automatic zero tracking)
Property Content Note
Command string ZTR
No. of parameters 1
Parameter range P1=0/1 (0= Off, 1= On)
Factory default 0 disabled
Reaction time < 15 ms
Password protection Yes
Relevant to legal for
trade
Yes
Parameter backup With command TDD1
Input Master ZTR(P1); No response
Query Master ZTR?;
Response WE P1crlf P1 = 1 character
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
Automatic zero tracking occurs with gross or net measured values < 0.5 d in the range ±2 %
of the nominal (rated) value of the scale (NOV
). The maximum adjustment speed is 0.5d/s at
scale standstill. The unit d (digit) relates to the nominal (rated) value (NOV
) and the
increment RSN
.
ZTR