WE2107 - Communication commands
I1820-1.3 en HBM
Trade Counter
(Legal for trade (calibration) counter)
Property Content Note
Command string TCR?
No. of parameters
Parameter range
Factory default < 00100
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
YES
Parameter backup
Query Master TCR?;
Response WE Xxxxx crlf 6 characters
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
This counter cannot be reset and marks the parameter changes of commands relevant to
legal for trade applications (see command LFT
). The maximum count is 65535. When this
count is reached, the counter stops data output MSV?
; then only outputs overflow values.
This status can only be removed at the factory.
It is only possible to read out the calibration counter.
TCR