WE2107 - Communication commands
34
I1820-1.3 enHBM
G‐Correction Factor
(G‐factor correction, calibration location)
Property Content Note
Command string ACA
No. of parameters 1
Parameter range P1 = 97000 ... 99000
Factory default 98102
Reaction time <15 ms
Password protection Yes
Relevant to legal for
trade
Yes
Parameter backup With command TDD1;
Input Master ACA(P1); No response
Query Master ACA?;
Response WE P1crlf P1 = 5 characters
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The command is used to correct the effect of gravitational acceleration when the place where
the scale is adjusted is not the same as the place of installation and the gravitational
acceleration factors (g) are different.
Parameter description:
For a query, the value is output 6‐digit (e.g. 098102 CRLF).
Correction is disabled at ACA = ACU.
Calculating the internal correction factor:
GF +
ǒ
g * factoratplaceofadjustment(ACA)
Ǔ
ǒ
g * factoratplaceofinstallation(ACU)
Ǔ
The internal GF parameter is reset automatically (ACU: = ACA) when
S curve SZA / SFA is re‐measured
S curve LDW / LWT is re‐measured
Input example:
Place of adjustment (ACA) = Darmstadt ³ g = 9.81029
Place of installation (ACU) = Tokyo ³ g = 9.7977
GF = 1.001285
ACA