WE2107 - Communication commands
26
I1820-1.3 enHBM
ASCII measurement format (COF4):
ASCII output always contains 16 characters.
Character
1
Characters
2‐10
Character
11
Characters
12‐14
Character
15,16
G
N
Measured value
(sign, measured
value with decimal
point)
Blank g
kg
t
lbs
pcs
crlf
G = gross,
N = net
9 x `‐` , when out
side display range
for
LT > 0
For standstill
only, otherwise
3 blanks
End
label
The display range is defined as follows:
LFT=0: ‐160 % .... + 160 % (cannot be verified, industrial mode)
LFT=1: ‐20 d ... + NOV + 9 d (legal for trade, OIML, R76)
LFT=2: ‐2 % ... + 105 % (NOV) (legal for trade, NTEP)
NOV
is the output scaling (NOV w 100). The percentage figures relate to the NOV.
The d information relates to increment that is set (RSN
):
RSN = 2 ‐> 9 d = 18 digits