33
WE2107 - Communication commands
I1820-1.3 en HBM
Manual input of the nominal (rated) value via SZA/SFA:
1. Connect the transducer electronics to a calibration standard.
2. Use the SPW
command to enter your password.
3. Set legal for trade switch to LFT= 0
4. Reset the correction factor: ACU:= ACU
5. Reset the factory characteristic: SZA=0, SFA=200000,
6. Reset the user characteristic: CWT=100000, LDW=0, LWT=200000,
7. NOV=0, RSN=1, switch the linearization OFF (LIN
, LIM).
8. Set the ASF filter in such a way that the display is as smooth as possible.
9. Set the calibration standard to 0mV/V misalignment.
10.Use the command MSV
?; to determine the measured value. Note value1 for SZA.
11.Set the calibration standard to 2mV/V misalignment.
12.Use the command MSV?; to determine the measured value. Note value2 for SFA.
13.Enter the new user curve with SZA<value1>; followed by SFA<value2>;.
14.Save the new curve with TDD1;.
NOTE
The characteristic curve commands SZA and SFA should be entered or executed in the
following order: SZA followed by SFA. The input data is only offset if these two parameters
have been entered in pairs.
When the factory default curve is entered with SZA/SFA, this resets the user curve to the
default values LDW=0, LWT=200000, CWT=100000 and ACU:=ACA.
Numbers 1 ‐ 10 do not apply if the factory default curve can be re‐entered using parameters
that are already known.
With the command TDD0; the factory default settings will be activated.