WE2107 - Communication commands
I1820-1.3 en HBM
Tare
(Tare with the actual gross value)
Property Content Note
Command string TAR
No. of parameters ‐
Parameter range ‐
Factory default ‐
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
no
Parameter backup No data to back up
Input Master TAR; No response
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The command TAR; tares the current measured value. After tare, it switches back to the net
measured value (TAS0;). The current value is stored in tare memory (also see the TAV
command) and subtracted from the measured value and all subsequent measured values.
With legal for trade applications (LFT>0) tare is only allowed when standstill conditions
occurs.
Permissible tare range:
LFT=0: " 100 % of NOV
LFT>0: 0… NOV
A TAR?; query is not permitted.
The stored tare value can be read out with TAV?. At Power OFF, the tare value is lost unless
it is saved with TDD1;.
TAR