WE2107 - Communication commands
56
I1820-1.3 enHBM
Filter mode
(Filter mode selection)
Property Content Note
Command string FMD
No. of parameters 1
Parameter range P1=0,1, 3, 4, 5
Factory default 0
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
No
Parameter backup With command TDD1
Input Master FMD(P1); No response
Query Master FMD?;
Response WE P1crlf P1 = 1 character
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
FMD0: normal behavior (as described with ASF
)
FMD1: fast settling mode
FMD2, 3, 4: Animal filter (with different settling times),
Recommendation: ASF should be set to 5.
FMD