83
WE2107 - Communication commands
I1820-1.3 en HBM
3.8 Commands for legal for trade applications
The commands are used to monitor parameter changes made to parameters Relevant to
legal for trade via the standard counter:
D Calibration switch LFT
D Calibration counter TCR?
The parameters Relevant to legal for trade are:
SZA
, SFA, ACA, ACU, CWT, LDW, LWT, NOV, RSN, MRA, MRB, MDT, ENU, DPT, LIN, LIM,
ZSE
, ZTR, LFT, TCR
If the calibration switch is set to LFT>0, it is not possible to make changes to parameters
Relevant to legal for trade. Before changing parameters relevant to legal for trade,
activate the password (DPW
, SPW) and set LFT to zero. Every LFT change increases the
calibration counter (TCR) that cannot be reset by 1.
After calibration, set LFT to a value greater than zero. Then read out the calibration counter
and make a note on the identification label of the scale.