WE2107 - Communication commands
I1820-1.3 en HBM
Multi Range Mode 2
(3‐range weighing display)
Property Content Note
Command string MRB
No. of parameters 1
Parameter range P1=0...NOV (99999) 0=disabled
Factory default 0
Reaction time <15 ms
Password protection Yes
Relevant to legal for
trade
Yes
Parameter backup With command TDD1
Input Master MRB(P1); No response
Query Master MRB?;
Response WE P1crlf P1 = 6 characters
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The command is used to set up the scale as a one, two or three ‐range scale. At MRA0;
2‐range mode is deactivated. With MRA< MRB ... NOV, it is possible to specify the
changeover point between ranges 2 and 3. If 3‐range mode is enabled, the set RSN
increment is valid for range 2. The increment for range 3 is then automatically the next
increment:
Examples (0 < MRA < MRB < NOV):
RSN=2: Range 1 with increment 2, range 2 with increment 5, range 3 with increment 10,
RSN=5: Range 1with increment 5, range 2 with increment 10, range 3 with increment 20
The display switches back to the increment of range 1 if the scale is unloaded.
MRB