WE2107 - Communication commands
94
I1820-1.3 enHBM
External display end character 2
(Protocol external display definition of the end character 2)
Property Content Note
Command string ED2
No. of parameters 1
Parameter range 0 … 31
Factory default 10
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
No
Parameter backup With command TDD1
Input Master ED2(P1); No response
Query Master ED2?;
Response WE xx crlf 2 characters
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
This command defines the second end character in the output string.
ED2=0: no end character is defined (the telegram is 1 byte shorter)
ED2=1…31 an end character is defined ( reference ASCII‐chart PC437)
Examples:
ED1=3: end character is ETX (=03hex)
ED1=10: end character is an LF (=0a hex)