145
WE2107 - Communication commands
I1820-1.3 en HBM
4.7 Setting a parameter in all the connected WE's
If the measurement query is executed properly, the parameters can be set in all the WEs
connected at the bus. Proceed as follows:
1. Select all the WEs by entering the broadcast command S98;
All the WEs execute the command, but do not send a response.
2. Enter the required parameters.
3. Use the command TDD1; to store the parameter power fail safe in the EEPROM.
4. Use Sii; to select the next WE, to read control parameters, for example.
This sequence can also be used when taring with the TAR; command, for example, or for
changing over between gross and net output.