69
WE2107 - Communication commands
I1820-1.3 en HBM
Query: TAV?;
The content of the tare memory is output. The tare value is converted to the NOV value.
With legal for trade applications (LFT>0), the tare range is restricted to 0...100 % of NOV.
permissible tare range :
LFT=0: ± 100 % of NOV
LFT>0: 0 NOV
Example:
NOV3000; (scale scaling)
TAS1; (gross output enabled)
MSV?; 1500 LF (measured value at 50% = max. capacity
of the scale)
TAR; (tare and select net output
TAV?; 1500 LF (query tare value)
MSV?; 0 LF (net measured value)
TAS?; 0 LF (net is enabled)
TAS1; (select gross)
MSV?; 3000 LF (measured value at 100 % = nominal (rated)
load of the scale)
TAV?; 1500 LF (query tare value, unchanged)