WE2107 - Communication commands
116
I1820-1.3 enHBM
Parameter P2 defines the function:
Digital input 1:
P2 code
function with SFU = 0
0 No function
1 Tare
2 Hard copy (selected with command PRT)
1)
3 Scale tilted (³ display tilt instead of measured values)
2)
P2 code function with SFU>0 (filling / dosing)
0 No function
1 Stop filling/dosing (BRK)
2 Hard copy (selected with command PRT)
1)
3 Scale tilted (³ display tilt instead of measured values)
2)
Digital input 2:
P2 code
Function with SFU = 0
0 No function
1 Tare
2 Hard copy (selected with command PRT)
1)
3 Lock parameters (only display of parameters in the parameter menu, a
change is not possible)
P2 code Function with SFU>0 (filling / dosing)
0 No function
1 Start portion weighing (RUN)
2 Hard copy (selected with command PRT)
1)
3 Lock parameters (only display of parameters in the parameter menu, a
change is not possible)
1)
before the use of this function, the print protocol has to be selected in the parameter menu Print\Result\PRT or
by the command PRT.
2)
the reaction time for the input TILT can be defined by the command TDL.