77
WE2107 - Communication commands
I1820-1.3 en HBM
With parameter input, the changed settings are initially only saved in working memory
(RAM), so they are not stored power fail safe. Use the command TDD1 to store the settings
that you have changed in working memory power fail safe in the EEPROM.
TDD0, restore the factory default:
This command restore the factory default parameters:
D SZA
, SFA (0…2 mV/V factory characteristic curve)
Unchanged parameters:
D Commands for the communication: BDR
, ADR, COF,
D Commands for the real time clock : TDT
,TME,TMM,
D Commands for the external display: EDP
, EDS, ED1, ED2, EDC
D Commands for the print setup: ESC,PES,
PID
,PLB,PLE,PRT,PST
Commands for setup buttons, digital inputs:
D BFL
, BFS, FIN, MAL, TDL
All other parameters are set to the factory default values as we described in this manual.
NOTE
After TDD0 the scale has to be adjusted again. The legal for trade switch is set to OFF. The
calibration counter is incremented by one.