89
WE2107 - Communication commands
I1820-1.3 en HBM
P1=1 / 3, actual Gross or Net value :
Byte
Content
1 Start character (EDS),
if EDS=0 no start char. will send out
2 Sign of the measured value
3 ‐ 9 P1=1: Gross or net value (with decimal point)
P1=3: Filling result (with decimal point)
1)
10 Empty space
11 ‐ 14 Unit (`kg ` ; `t ` ; `g ` ; `lbs ` ; `pcs ` ), if standstill otherwise empty spaces
15 Empty space
16 – 19 `G ` =Gross value; `Net ` = Net value or
`N PT' = Net value with usage of pre tare value
20 End character 1 (ED1),
if ED1=0 no char. will send out
21 End character 2 (ED2),
if ED2=0 no char. will send out
22 Check sum character (EDC),
if EDC=0 no char. will send out
1)
if an error occurs than the value is `Err xx'