b. Remove the incubator cover.
c. Unscrew 5 screws on upper incubator and remove
the incubator.
d. Wipe off 13 pressure plates on the underside of the incubator
with soft gauze-like cloth or cotton swab moistened and
tightly wrung out with sterilizing ethyl alcohol or isopropyl
alcohol.
e. Dry the pressure plates by facing the incubator backside
(pressure plate surface) up.
IMPORTANT
Do not touch the surface of the pressure plates with bare hands.
Otherwise, adverse effects on test results may occur.
f. Clean the incubator cells in the same way.
g. Unscrew 2 screws fixing the reference plate and remove the
plate from the body.
Thumbscrew
Pressure plate
Incubator cover
Reference plate
Thumbscrew