Operation
Item Designation Description/function
1
‘Manual mode’
tab
Chapter 6.11 ‘Manual mode’
on page 54
Here, you can make settings for manual mode.
2
Start delay
‘Start delay’
on page 57
Here, you can dene the date and time for the starting
point of manual mode. If the start delay is deactivated,
manual mode is started without a time delay, directly after
activating manual mode.
3 Temperature drop (for incubators
with a cooling unit only)
‘Cre-
ating a temperature drop’
on page 60
Temperature drops can be activated, deactivated or set.
4
[Go back]
button You can return to the main screen via the
[Go back]
button.
5
[Start]
button By pressing the
[Start]
button, manual mode is activated
with the corresponding settings, taking the start delay into
account.
6
PIN protection
‘PIN protection’
on page 64
To protect the settings from being changed, you can acti-
vate/deactivate the PIN protection here. After starting
manual mode, the
‘Admin PIN’
must be entered rst before
changes are possible.
7 Control contact Here, you can activate/deactivate the control contact.
This option is only available if the incubator is equipped
with the Switchboar
d option
Chapter 6.13.6.1 ‘Switch-
board option’ on page 102
or passive dehumidication
option
Chapter 6.13.6.2 ‘Passive dehumidication
option’ on page 104
and the control contact is activated in
the device settings
Chapter 6.13.6 ‘Control contact
(option)’ on page 101
.
8 Actual interior temperature Indicates the current, unregulated actual temperature of the
interior.
9
Set temperature
‘Set tempera-
ture’ on page 65
You can change the temperature setpoint in steps of 0.1
with the
[Plus]
and
[Minus]
buttons. By tapping the temper-
ature display, you can make inputs via a keypad on the
screen.
When manual mode is started, just like when pro-
gram mode is started, the setting of the temperature
guard is checked
Chapter 6.13.5 ‘Temperature
guard’ on page 98.
If the temperature setpoint set for manual mode
does not agree with the temperature guard setting, a
warning message will appear. If the warning mes-
sage is conrmed, manual mode starts despite the
contradiction. Check and correct the values of the
temperature guard and/or the temperature setpoint.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals