4.3 Integration Value Observation
61
4
Chapter 4 Viewing Measurement Values
After pressing , the instrument waits until the specified clock start time. Integration then begins and
continues until the specified clock stop time.
If auto-saving is enabled, integration values are saved to CF card at the specified start and stop times. If
an interval time is also specified, total integration values up to that point are saved after each interval.
Clock-Controlled Integration
Setting Procedure
Display the
[Time] page
Select [Real time]
to enable (set to
[ON])
Select a [Start time] and
[Stop time] digit
(automatically starts and stops
at the specified clock times)
During wait: blinks green
At the start time: lights green
To abort automatic control (while waiting):
Press again.
Use the F keys to set the timer
• Clock control settings are in 1-minute units.
• Clock setting years are AD (Christian Era), and 24-hour time (e.g., December 6th 2009,
10:16 PM appears as 2009-12-06 22:16)
• If a specified clock time has already passed, clock control is considered to be disabled
(OFF).
• When integration is interrupted during clock timing control, clock control is disabled (OFF).
• If clock timing control is set longer than the timer setting, integration starts at the clock start
time and stops when the timer has expired (the clock stop time is ignored).
• Integration stops after 9999 hours, 59 minutes, and 59 seconds if the time between clock
start and stop times is set longer than that.
• Setting time upper limits are as follows:
Start Time 2077-12-31 23:59
Stop Time 2079-12-31 23:59