Hioki 3390
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10.3
Setting Specifications
160
10.3 Setting Specifications
1. Input Settings
Wiring modes
Synchronization source U1 to U4, I1 to I4, Ext (when channel B is set for pulse input with a motor evaluation option installed)
DC (50 ms/100 ms) @Selectable on all wiring systems
Voltage range AUTO, 1500 V, 600 V, 300 V, 150 V, 60 V, 30 V, 15 V
Voltage rectification method RMS / MEAN (voltage value used to calculate apparent and reactive power, and power factor)
Current range AUTO, 20 A, 8 A, 4 A, 2 A (20 A with 9272-10)
AUTO, 20 A, 8 A, 4 A, 2 A, 0.8 A, 0.4 A (20 A with 9277)
AUTO, 200 A, 80 A, 40 A, 20 A, 8 A, 4 A (200 A with 9272-10, 9278 and CT6863)
AUTO, 50 A, 20 A, 10 A, 5 A, 2 A, 1 A (CT6862)
AUTO, 500 A, 200 A, 100 A, 50 A, 20 A, 10 A (9279, 9709)
Current rectification method RMS/ MEAN (current value used to calculate apparent and reactive power, and power factor)
VT(PT) ratio OFF/ 0.01 to 9999.99(setting not available if VT×CT ratio exceeds 1.0E+06)
CTratio
OFF/ 0.01 to 9999.99(setting not available if VT×CT ratio exceeds 1.0E+06)
LPF
OFF, 500 Hz, 5 kHz, 100 kHz
Lower limit measurement frequency 0.5 Hz, 1 Hz, 2 Hz, 5 Hz, 10 Hz, 20 Hz
Frequency measurement Select U or I for f1, f2, f3, and f4
Integration mode RMS/ DC
CH1 CH2 CH3 CH4
Mode 1 1P2W 1P2W 1P2W 1P2W
Mode 2 1P3W 1P2W 1P2W
Mode 3 3P3W2M 1P2W 1P2W
Mode 4 1P3W 1P3W
Mode 5 3P3W2M 1P3W
Mode 6 3P3W2M 3P3W2M
Mode 7 3P3W3M 1P2W
Mode 8 3P4W 1P2W
2. Calculation and Recording Settings
Average OFF, FAST, MID, SLOW
Interval OFF, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 5 s, 10 s, 15 s, 30 s,
1 min, 5 min, 10 min, 15 min, 30 min, 60 min
Timer control Elapsed time / Real-time clock
Timer:..... Off, 10 s to 9999:59:59 [hhhh:mm:ss] (in 1 s units)
Real-time clock:OFF, start and stop times (YMD-hms, in 1 min units)
Zero suppression OFF, 0.1%/ 0.5%f.s.
Zero-crossing filter Off, mild or strong
Auto-ranging span Wide or narrow
Efficiency calculations Three items (select from all active power values) η=100 × |Pout|/|Pin|
Loss calculations Three items (select from all active power values) Loss=|Pin| - |Pout|
Δ- Y Conversion OFF/ ON
3. Harmonic Settings
Harmonic U1 to U4, I1 to I4, Ext (when channel B is set for pulse input with a motor evaluation option installed)
DC (50 ms/100 ms) Settings common to all channels
TTHD calculation THD-F/ THD-R
4. Noise Analysis Settings
Measurement cannels Select one of channels 1 to 4
Windows Rectangular, Hanning, flat-top
Lower limit noise frequency 0 kHz to 10 kHz
5. D/A Output Settings (with D/A output option Model 9792 or 9793)
Waveform output Off or On (See separate specifications for Model 9793 when enabled [ON])
Output items Select one basic measurement item for each output channel.
Only selectable for channels 9 to 16 when waveform output is enabled [ON]
(Channels 1 to 8 provide waveform output only)
Full-scale frequency integration 100 Hz, 500 Hz, 1 kHz, 5 kHz (same as max. measurement frequency setting for motor)
Full scale 1/10, 1/2, 1/1, 5, 10, 50, 100, 500, 1000, 5000, 10000 × range