l
SYSTem:DUMP:PRINter?
Causes a dump of the display contents to be made to the HP-IB.
l
SYSTem:ERRor?
Returns the next message in the error queue. The format of the
response is :
<error
number>,<error
string>
where the error number is as shown in the “Error Messages” section
and error string is :
“<Generic HP-SL
string>;<More
specific information>”
An
example response to SYST:ERR? is
-23,"NUMERIC 0VERFLOW;YOUPUT IN A NUMBERTOO BIG"
l
SYSTem:LANGuage
SCPIlCIILlCOMPatible
Causes the swept signal generator to perform a language switch to
another language system.
l
SYSTem:MMHead:SELect:AUTO
ONIOFFIOll
l
SYSTem:MMHead:SELect:AUTO?
Sets and queries automatic selection of the millimeter source
module interface connector. If the swept signal generator finds
a source module connected at both the front and the rear
connectors, the source module at the front connector is selected.
Programming a specific state via SYSTem:MMHead:SELect sets
SYSTem:MMHead:SELect:AUTO to OFF. *RST value is 1.
l
SYSTem:MMHead:SELect
FRONtlREARlNONE
a
SYSTem:MMHead:SELect?
Sets and queries the active millimeter source module interface.
Programming a specific state
(FRONt
[REAR/NONE) sets
SYSTem:MMHead:SELect:AUTO to OFF. Programming a specific
state will cause the instrument to examine the selected interface to
determine the type of source module connected. The instrument
frequency limits and multiplier will be altered accordingly. However,
the leveling point is not changed (see POWer:ALC:SOURce).
l
SYSTem:PRESet[:EXECutel
Sets the swept signal generator to its “local operation” state. This
is the same as pressing the front panel “green” key. There is no
corresponding query.
l
SYSTem:PRESet:SAVE
Saves the present state so it can be used whenever the command
SYSTem:PRESet[:EXECute] is executed or the front panel “green"
key is pressed.
l
SYSTem:PRESet:TYPE
FACTorylUSER
l
SYSTem:PRESet:TYPE?
Operating and Programming Reference S-53