Page 10 - CEAST 9340 - Instructions for Use and Maintenance 7510.000MN1r ed. 1 rev. 1
HAZARD IDENTIFICATION
Nitrogen is colourless, odourless, non-flammable, non-irritating and non-toxic gas.
It is classified as a Simple Asphyxiant under OSHA regulations.
Can cause suffocation by reducing oxygen available for breathing.
Breathing very high concentration of vapour can cause dizziness, shortness of
breath, unconsciousness, or even death.
There are no currently Occupational Exposure Limit Values for this component
but it is recommended to maintain oxygen levels at or above 19,5%.
Use only in a well ventilated place.
Local exhaust system of 1200 m
3
/hour is recommended
to prevent high nitrogen concentration when use the CEAST 9340
equipped with the environmental chamber.
Avoid direct inhalation of undiluted gas.