9.3 CLEANING AND DRYING METHODS
Authorized Personnel: OPERATOR
- Remove the fragments of the specimens from the chamber using an industrial type
vacuum cleaner.
- Clean the outer painted surfaces of the instrument with soft rags and non corrosive,
harmless detergents. In any case the safety rules foreseen by the manufacturer of
the cleaning substances must absolutely be adhered to.
Every day, at the end of low temperature tests, heat the environmental cham-
ber to approx. 70 °C for at least two hours, leaving the door open.
Dry the surfaces of the stand installed in the chamber (i.e.: adjustable height
stand, fixed height stand, pivoting loader, automatic specimens feeder and
related clamping ring) with a soft, dry rags. After drying, lubricate the sur-
faces of these parts with silicone oil spray or normal lube oil.
The purpose of above-mentioned operations is to eliminate condensate from
the environmental chamber and to avoid the formation of rust.