CEAST 9340 - Instructions for Use and Maintenance 7510.000MN1r ed. 1 rev. 1 - Page 115
X
X
X
a) Turn on the instrument and wait that it performs the firmare autocheck.
b) If the display visualizes an error status:
- Press the “Enter” key. This command allows to repair the error condition of firm-
ware.
- Verify the loader and firmware version from the menu <\Utility\FW Version>.
- Go in the sub-menu <\Service\Modify Configuration> and <\Service\Modify Cal.
Par.> and compare all the calibration parameters with those previously noted in
the notebook (see Note, para. 9.4.2).
c) If the display do not visualizes an error status, proceed with the use of the instrument.
9.4.5 INSTRUMENT FIRMWARE CHECK
X
X
X
a) Close the window (CLOSE button) and the loading program (EXIT option).
b) Turn off the PC and the instrument.
c) Disconnect the serial cable from the instrument and PC.
9.4.4 DISCONNECTION