APPENDIX F - OUTPUT LIST
OUTPUT LIST (1 of 2)
Output Meaning IF
ACTIVE (logic
state "1" writed
by CPU -- LED
on board ON--
Output driven
toward positive
reference)
Power Supply Actuator Notes
1 POW_OUT_1
(M5-12
LED D 36)
Open the shutter ALIM_1
24V_ MOVE
EV chamber
shutter (bistable)
Opening command
Controllable on
CPU through GPIO
PP4 deactivating
the PWM4
(Port P -- bit4)
2 POW_OUT_2
(M5-14
LED D 37)
Close the shutter ALIM_1
24V_ MOVE
EV chamber
shutter (bistable)
Closing command
Controllable on
CPU through GPIO
PP5 deactivating
the PWM5
(Port P -- bit5)
3 POW_OUT_3
(M5-16
LED D 38)
Release the motor
electromechanical
brake
ALIM_1
24V_ MOVE
Triac The motor is of
self-braking type:
brake command
absent --> brake
active
4 POW_OUT_4
(M5-18
LED D 42)
Release the mass ALIM_1
24V_ MOVE
EV Releasing
(monostable)
When powered it
releases; normally
hokeed
5 POW_OUT_5
(M5-20
LED D 43)
Antirebound active
(Up)
ALIM_1
24V_ MOVE
EV Antirebound
(bistable)
Up command
6 POW_OUT_6
(M5-22
LED D 44)
Antirebound
inactive
(Down)
ALIM_1
24V_ MOVE
EV Antirebound
(bistable)
Down command
7 POW_OUT_7
(M5-24
LED D 47)
Specimens
clamping Up
POW_COM_7
24V_DOOR
EV Clamp
(bistable)
Up command
8 POW_OUT_8
(M5-26
LED D 48)
Specimens
clamping Down
POW_COM_8
24V_DOOR
EV Clamp
(bistable)
Down command

Table of Contents