6. INSTALLATION
6.1 SPACE AND ENVIRONMENT REQUIREMENTS
6.1.1 SPACE
6.1.2 ENVIRONMENT
6.1.3 SERVICES
Installation of the instrument must be performed according to the instructions, by trained
qualified personnel, Mechanics or Electricians as specified according to the needs.
The instrument will work according to the technical parameters foreseen if it is correctly
placed, stably, in a laboratory on the floor that is able to hold the instrument weight.
Before installing the instrument, determine the correct place to position it. Remember
that it is necessary to consider the following requirements: space restraints, absence of
vibrations, temperature, humidity and atmosphere (not corrosive, not polluted, not ex-
plosive).
The spatial dimensions of the instrument are indicated in paragraph 1.2.1. Remember
to leave adequate space all around the instrument (about 1 meter) and on top of it
(about 1 meter) to allow the access of maintenance personnell.
The instrument must be installed in a controlled temperature and humidity laboratory. If
this is not possible, it should be located in a clean environment where the temperature
range is between 15°C and 35°C with relative humidity (RH%) ≤ 85 % (without conden-
sation).
The location determined for the installation of the instrument must have an electrical socket
according to the information on the ID plate of the instrument (see chapter 1) and a pneu-
matic air supply line (see paragraph 1.2) able to delivery compressed air with pressure 7 to
10 bar.