EasyManuals Logo

Integra Cusa Excel User Guide

Integra Cusa Excel
189 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
General Troubleshooting
Troubleshooting
the CUSA EXcel System
CUSA Excel Ultrasonic Surgical Aspirator System User’s Guide 13-9
Little or no vibration or fragmentation
(Amplitude LEDs may illuminate well
below setpoint)
Excessive tip/tissue pressure (stalling tip) 1. Verify that console settings are at desired
levels.
2. Remove the tip from contact with tissue,
then reactivate vibration.
Note: When first activating vibration, make
sure that there is no tip/tissue pressure.
Blocked tip 1. Verify that the manifold suction tube
connects properly to the nosecone by
disconnecting, then reconnecting, the
tubing and its clip.
2. Use the tip cleaner to dislodge the
blockage in the tip. Continue to push the
tip cleaner into the tip until the cleaner
becomes visible in the tubing.
3. When you have cleared the blockage,
verify again that the manifold suction tube
connects properly to the nosecone.
4. If the tip remains blocked, disconnect the
manifold suction tube connector from the
nosecone. Use the tip cleaner to dislodge
blockage from the tubing, then reconnect
the tubing to the nosecone.
5. Immerse the handpiece tip in sterile
irrigation fluid briefly. This may clear the
suction tubing.
Incorrect flue installed Install the correct flue.
Condition Causes Recommended Action

Table of Contents

Other manuals for Integra Cusa Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Integra Cusa Excel and is the answer not in the manual?

Integra Cusa Excel Specifications

General IconGeneral
BrandIntegra
ModelCusa Excel
CategoryMedical Equipment
LanguageEnglish

Related product manuals