200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 2-15
2. Installation Procedures
20. Run a patient barcoded sample tube with diluent on the AX-4280 and
verify the rack transfers to the iQ 200 without error and the chemistry
results attach to the microscopy results.
21. If the chemistry results do not cross over to the worklist, check the LIS
output on the AX-4280 for correct setup:
i. Enter “-4280-
ii. Enter 1” for user setup
iii. Press MENU
iv. Press 2” for Online
v. Press ENTER” until “RS232C Normal” is displayed
vi. Change “RS232C Normal” to ”ON” by pressing “-“, then press
ENTER
vii. Change “RS232C STAT” to “ON” by pressing “-“, then press ENTER
viii. Change “RS232C Ctrl” to “ON” by pressing “-“, then press ENTER
ix. SAVE? Yes (1) No (0) Press “1” to save
x. Press ESC”, then press “STOP”, AX-4280 will initialize
22. If the chemistry cross over but do not attach to the microscopy, make
sure the barcode reader is setup correctly and reads the label the
same on both units. The label ID’s should match in order to attach the
chemistry and microscopy results together.