Iris Diagnostics, a Division of Iris International, Inc.
iQ
®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 2-14
2. Installation Procedures
16. Load the AX Connectivity CD PN 700-3106 on the Analysis
Processor.
17. From the Main Menu screen of the iQ, click on the Settings button.
18. Click on Interfaces.
19. Verify that the “Enable direct connection” checkbox with the AX-
4280 is selected.