384 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
error message Initialization Error (Firmware CRC wrong: crc1, stored: crc2!)
description The checksum stored in the firmware does not match the calculated checksum
error value crc1 calculated checksum
crc2 checksum read from caldata
action Reload the firmware of the module. If the failure persists, please return the unit to your local
service center for repair
error message Initialization Error (HwAccessInit, error value)
description Error setting DAC
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (InitDacs, error value)
description Error setting DAC
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (InstSettingInit, error value)
description Error creating semaphore
error value 2 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (LaserEEPROMInit, error value)
description Error reading laser EEPROM
error value 1 (not significant)
action Please return the unit to your local service center for repair

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A and is the answer not in the manual?

Keysight Technologies 8163A Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A
CategoryMultimeter
LanguageEnglish

Related product manuals