8163A/B, 8164A/B, 8166A/B User’s Guide 385
Error Messages 14
error message Initialization Error (MainEEPROMInit, error value)
description Error reading main EEPROM
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (MotorSelfCali failed, error value)
description Motor self calibration failed
error value -1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (MotorZero, error value)
description DSP motor zero not finished in time
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (OptZero, error value)
description Optical zero disabled in caldata
error value -1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (OptZero Opc, error value)
description Timeout while waiting for laser diode (16) and thermopositioner (64) to settle while preparing
optical zero measurement
error value 16 waiting for laser diode
64 waiting for thermopositioner
action Reboot the instrument. If the failure persists, please return the unit to your local service center
for repair

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A and is the answer not in the manual?

Keysight Technologies 8163A Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A
CategoryMultimeter
LanguageEnglish

Related product manuals