200 Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
System Key Group
Timestamp Displays the following softkeys to clear timestamp.
Clear Clears the timestamp. Press this softkey to display the
confirmation dialog box. Press the OK softkey to clear the
timestamp, or the Cancel/Local key to cancel.
Auto CLR Sets the automatic clear of the timestamp. The present setting
is displayed on the softkey label. Press the key to switch the
setting.
OFF Disables the automatic clear of the
timestamp.
ON Enables the automatic clear of the
timestamp.
Start-up Displays the System Start-up dialog box which provides the following setup
parameters.
Power-on State Power-on state, RST, #1 (RCL0), #2 (RCL1), #3 (RCL2), #4
(RCL3), or #5 (RCL4)
The power-on state can be selected from the factory default
reset condition (RST) and user conditions #1 (RCL0), #2 (RCL1),
#3 (RCL2), #4 (RCL3), and #5 (RCL4). The user conditions are
defined by the *SAV commands, or the destinations where the
system setting was saved by the Save key.
#1 (RCL0): Destination #1 or *SAV0 command
#2 (RCL1): Destination #2 or *SAV1 command
#3 (RCL2): Destination #3 or *SAV2 command
#4 (RCL3): Destination #4 or *SAV3 command
#5 (RCL4): Destination #5 or *SAV4 command
Power-on
Program Power-on program execution, ON or OFF
If this parameter is ON, the power-on program is automatically
executed when the B2980 is turned on.
The power-on program can be defined by using the
:PROG:PON:COPY command while the B2980 is in the remote
mode.