190
TOS3200
A List of Messages
SYSTem subsystem
TRIGger subsystem
SCPI Command Setting
Default
Respon
se
Affec
ted
Description R/W *
Program header
Parameter Unit
SYST
:BEEP:VOL
:FAIL numeric 0 to 10 3 NR3
Sets the volume of the buzzer that
sounds when the judgment is FAIL.
R/W 3
:PASS numeric 0 to 10 3 NR3
Sets the volume of the buzzer that
sounds when the judgment is PASS.
R/W 3
:CONF
:CONV numeric
0,
80.0 to 300.0
V0NR3
Sets the supply voltage used to convert
to current.
Specify zero when not converting to
current.
R/W 3
:LBR bool ON NR1
Sets whether or not to shut off the
power supply to the EUT during the
interval time.
R/W 3
:MMOD char NORM | MAX NORM char Yes
Sets whether to hold the maximum
measured current.
R/W 3
:PHOL numeric
0.2 to 10.0,
9.9E37
S2NR3
Sets the time to hold the PASS setting.
9.9E37 is hold.
R/W 3
:SELV numeric 0, 10 to 99 V 0 NR3
Sets the safety extra low voltage
(SELV).
Specify zero to turn the SELV function
off.
R/W 3
:TRAC bool OFF NR1
Turns on/off the communication error
trace function.
R/W 3
:DATE
NR1 2002 to 2099 NR1 Sets the year.
R/W 1NR1 1 to 12 NR1 Sets the month.
NR1 1 to 31 NR1 Sets the day.
:ERR[:NEXT] Queries the error and event information. R 1
:KLOCK bool OFF NR1 Turns on/off the panel operation lock. R/W 1
:LOC
Sets to local mode
(RS232C and USB only).
W2
:OPT char Queries the options. R 3
:PROT:CLE Releases the protection status. W
:REM
Enables remote mode and locks all key
and switches besides the LOCAL key
and STOP switch
(RS232C and USB only).
W2
:RWL
Enables remote mode and locks all key
and switches besides STOP switch
(RS232C and USB only).
W2
:TIME
NR1 0 to 23 NR1 Sets the hour.
R/W 1NR1 0 to 59 NR1 Sets the minute.
NR1 0 to 59 NR1 Sets the second.
:VERS 1999.0
Queries the SCPI specification version
with which the TOS3200 complies.
R1
SCPI Command Setting
Default
Respon
se
Affec
ted
Description R/W *
Program header
Parameter Unit
ABOR Aborts the test. W 1
INIT[:IMM]
[:SEQ[1]]
Starts the test (trigger function).
W1
:NAME char TEST W1