Gocator Snapshot Sensors: User Manual
Gocator Web Interface • 94
If Video mode is selected, tools will not be available in the Measure panel.
4. Expand the Alignment panel by clicking on the panel header or the button.
5. Ensure that all sensors have a clear view of the target surface.
Remove any irregular objects from the sensor's field of view that might interfere with alignment.
To perform stationary alignment:
1. In the Alignment panel, select Stationary as the Type.
2. Clear the previous alignment if present.
Press the Clear Alignment button to remove an existing alignment.s
3. Select an alignment Target.
4. Place the target under the sensor
5. Click the Align button.
The sensors will start, and the alignment process will take place. Alignment is performed
simultaneously for all sensors. If the sensors do not align, check and adjust the exposure settings (page
83).
Alignment uses the exposure defined for single exposure mode, regardless of the
current exposure mode.
6. Check the alignment results under Transformation in the Active Area tab in the Sensor panel.
Clearing Alignment
Alignment can be cleared to revert the sensor to sensor coordinates.
To clear alignment:
1. Go to the Scan page.
2. Choose Surface mode in the Scan Mode panel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LMI Technologies Gocator 3210 and is the answer not in the manual?

LMI Technologies Gocator 3210 Specifications

General IconGeneral
BrandLMI Technologies
ModelGocator 3210
CategoryAccessories
LanguageEnglish

Related product manuals