Page 20
Functional Principle, Technical Data
optoNCDT 2300
Model ILD2300- 2 5 10 20 50 100 200 300
Measuring range
1
mm
(inches)
2 / 2
(.08 / .08)
5 / 2
(.20 / .08)
10 / 5
(.39 / .20)
20 / 10
(.79 / .39)
50 / 25
(1.97 / .99)
100 / 50
(3.94 / 1.97)
200 / 100
(1.57 / 3.94)
300 / 150
(11.81 / 5.9)
Start of measuring range
1
mm
(inches)
24 / 24
(.94 / .94)
24 / 24
(.94 / .94)
30 / 35
(1.18 / 1.38)
40 / 50
(1.57 / 1.97)
45 / 70
(1.77 / 2.76)
70 / 120
(2.76 / 4.72)
130 / 230
(5.12 / 9.06)
200 / 350
(1.57 / 13.78)
Midrange
1
mm
(inches)
25 / 25
(.98 / .98)
26.5 / 25
(1.04 / .98)
35 / 37.5
(1.38 / 1.48)
50 / 55
(1.97 / 2.17)
70 / 82.5
(2.76 / 3.25)
120 / 145
(4.72 / 5.71)
230 / 280
(9.06 / 11.02)
350 / 425
(13.78 / 16.73)
End of measuring range
1
mm
(inches)
26 / 26
(1.02 / 1.02)
29 / 26
(1.14 / 1.02)
40 / 40
(1.57 / 1.57)
60 / 60
(2.36 / 2.36)
95 / 95
(3.74 / 3.74)
170 / 170
(6.69 / 6.69)
330 / 330
(13.0 / 13.0)
500 / 500
(19.69 / 19.69)
Measuring rate
7 adjustable stages: 49.14 kHz
2
/ 30 kHz / 20 kHz / 10 kHz / 5 kHz / 2. kHz / 1.5 kHz
Linearity
µm < ± 0.6 < ± 1.5 < ± 2 < ± 4 < ± 10 < ± 20 < ± 60 < ± 90
% FSO < ± 0.03 < ± 0.02 < ± 0.03
Resolution
3
µm 0.03 0.08 0.15 0.3 0.8 1.5 3 4.5
Light spot diameter
(±10 %)
SMR, µm 55 x 85 70 x 80 75 x 85 140 x 200 255 x 350 350
1300
580 x 860
MR, µm 23 x 23 30 x 30 32 x 45 46 x 45 70 x 70 130 380 x 380
EMR, µm 35 x 85 70 x 80 110 x 160 140 x 200 255 x 350 350 470 x 530
Light source Semiconductor laser < 1 mW, 670 nm (red); Class 2 acc. to IEC 60825-1: 2014 / optional class 3R
Permissible ambient light 10,000 lx … 40,000 lx
Power supply 24 VDC (11 ... 30 V); < 3 W (24 V)
Digital interface
RS422 (16 bit) / Ethernet / EtherCAT / PROFINET
4
/ EtherNet/IP
4
Analog output
4
4 … 20 mA / 0 … 5 V / 0 … 10 V / ± 5 V / ± 10 V
Temperature range Operation: 0 ... +50 °C (+32 °F up to +122 °F); storage: -20 ... +70 °C; (non-condensing)
Shock (DIN EN 60068-2-29) 15 g / 6 ms in 3 axis
Vibration (DIN EN 60068-2-6) 2 g / 20 … 500 Hz
Protection class (DIN EN 60529) IP65
Housing size, material S, die-cast zinc M, aluminum
Weight approx. 550 g (incl. pigtail)
FSO = Full Scale Output, SMR = Start of measuring range, MMR = Mid of measuring range, EMR = End of measuring range; The specified data ap-
ply to white, diffuse reflecting surfaces (Micro-Epsilon reference ceramic for ILD sensors)
1) Second value is valid for a measuring rate of 49.14 kHz 3) Measuring rate of 20 kHz
2) Measuring rate of 49.14 kHz with reduced measuring range 4) Connection via interface module (see accessories)