Page 87
Digital Interfaces
optoNCDT 2300
Displacement values, differential values, intensity
An intensity for each selected displacement, if selected, is transmitted.
Maximum 2 measurement values can be transmitted by direct reflection.
Displacement or differential value
Ethernet: The displacement values or the differential value were shown as a signed integer value (32 bit) with a resolution of 1 nm.
RS422: The displacement values or the differential value were shown with a resolution of 16 bit, while the value range is 18 bit. This
means, that even D16 and D17 stand in the H-byte, when through a denser medium with a refractive index >1, maximal 4, is mea-
sured through, see Chap. 8.3.
Intensity
Bit 31 ... 25 24 ... 14 13 ... 10 9 ... 0
Reserved
Maximum of the peak
(from the corrected signal)
Reserved
Raw-Intensity of the
peak
If the intensity is transmitted via the RS422, only bit 0 ... 9 is transmitted.
Status
17, 16 15 14 ... 7 6 5 4 3 2 1 0
Status LED
0 0 off
0 1 green
1 0 red
1 1 yellow
Trigger identifier
0 no trigger
1 trigger
Reserved Peak is lo-
cated be-
hind the
measu-
ring range
Peak is
in front
of the
measu-
ring range
(MR)
not all peaks
were calculated
(peaks are too
close together)
there are
fewer peaks
selected as
applicable
there is
no peak
present
Peak
ends to
late
Peak
starts to
early
If the status is transmitted via the RS422, only bit 0 ... 17 is transmitted. Bit 0 ... 6 is identical to the error codes RS422.
Trigger counter
Bit 31 30 29 ... 16 15, 14 13 ... 0
Identification of
the trigger
Reserved
Trigger event counter
(TriggEventCnt)
Reserved
Trigger measuring value counter
(TriggValueCnt)
The trigger counter can not be transferred with the RS422.