EasyManuals Logo

Mindray DPM 4 Operator's Manual

Mindray DPM 4
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #188 background imageLoading...
Page #188 background image
14-6
Figure 14-6 IBP Alarm Setup Menu
IBP alarm limits
Pressure Label Upper Limit (mmHg) Lower Limit (mmHg) Step (mmHg)
ART (lower limit+2) ~300 0~(upper limit-2) 1
PA (lower limit+2) ~120 -6~(upper limit-2) 1
CVP (lower limit+2) ~40 -10~(upper limit-2) 1
RAP (lower limit+2) ~40 -10~(upper limit-2) 1
LAP (lower limit+2) ~40 -10~(upper limit-2) 1
ICP (lower limit+2) ~40 -10~(upper limit-2) 1
P1 (lower limit+2) ~300 -50~(upper limit-2) 1
P2 (lower limit+2) ~300 -50~(upper limit-2) 1
The PR alarm limits are the same as those of the corresponding SpO
2
module.
SCALE ADJUST >> This option allows you to open the following menu and set
the positions of the high, reference and low scales for the
two waveforms.
Figure 14-7 IBP Scale Adjust Menu

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals