Command reference
R&S
®
ZNA
1272User Manual 1178.6462.02 ─ 20
Manual operation: See "Auto Length / Auto Length and Loss" on page 665
[SENSe:]CORRection:COLLect:PMETer:ID <PowerMeter>, <TestPort>
Selects an external power meter for the SMARTerCal and assigns it to an analyzer
port.
Note: The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Parameters:
<PowerMeter> Number of external power meter. The parameters UP, DOWN,
MIN, MAX are not available for this command.
Range: 1 to number of configured external generators
*RST: The power meter selection is not changed by a
reset of the analyzer.
<TestPort> Test port number. During power calibration, the power meter is
connected to this port.
*RST: 1 to the number of test ports
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
Manual operation: See "Power Meter" on page 531
[SENSe:]CORRection:IMEThod <InterpolationMethod>
Selects the algorithm for the interpolation of user system error corrections.
For the factory system error correction data, the analyzer always uses linear interpola-
tion.
Parameters:
<InterpolationMethod>LINear | HORDer
LINear
Linear interpolation
HORDer
Higher order (cubic spline) interpolation
Manual operation: See "Interpolation" on page 790
[SENSe<Ch>:]CORRection[:STATe] <Boolean>
Enables or disables the system error correction for channel <Ch>.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the correction.
*RST: ON
SCPI command reference