Command reference
R&S
®
ZNA
1280User Manual 1178.6462.02 ─ 20
Manual operation: See "CalUnit Con" on page 533
Test Port Port Assignment 1 Port Assignment 2
2 Cal Unit Port 1 -
3 Cal Unit Port 2 auto-detected
4 - auto-detected
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
<TestPort1>[, <TestPort2>]...
Similar logic as [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:
DEFine, but initially defines the port assignment only by its underlying test port set; the
connected calibration unit ports are auto-detected at the start of the corresponding
calibration sweep.
Suffix:
<Ch>
.
Number of the channel
<Asg> Number of the port assignment
Parameters:
<TestPort1> First test port number
<TestPort2> Second test port number ...
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1278
Manual operation: See "CalUnit Con" on page 533
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure <CalType>,
<Characterization>[, <CalName>]
Selects a calibration type and a cal unit characterization (cal kit file) for an automatic
calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalType> FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort | PFNPort | VUOSm | VPUosm |
CPUosm
Calibration Type
FNPort
Full n-port (UOSM) calibration
SCPI command reference