GUI reference
R&S
®
ZNA
533User Manual 1178.6462.02 ─ 20
CalUnit Con ← Cal. Settings
Opens the "Calibration Unit" dialog that allows you to define the port assignments for
an automatic calibration.
The R&S ZNA always proposes an optimum solution (minimum number of assign-
ments) that also minimizes the physical port reconnections required between calibra-
tion stages. For user-modified assignments, it provides assistive information indicating
insufficient, or redundant entries.
The test port connectors are automatically set according to the connector type of the
selected calibration unit port.
"Detect Port Assignment" starts a procedure by which the R&S ZNA (with a little help
from the attached calibration unit) auto-detects the connected ports. The automatic
assignment replaces the configured one.
In case auto-detection fails
an error report is shown as a warning dialog
the undetected port connections are marked with warning signs
the calibration can be invalid
"Default Port Assignment" restores the default port assignments.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?