Command reference
R&S
®
ZNA
1311User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
Parameters:
<ElecLength> Electrical length
Range: -1E+9 m to +1E+9 m.
Increment: 1 mm
*RST: 0 m
Default unit: m
Example:
CORR:EDEL2:ELEN 0.3
Define an electrical length of 30 cm for channel 1 and port no. 2.
CORR:EDEL2:DIST?; DIEL?
Query the values of the mechanical length and the permittivity at
port 2. The mechanical length is equal to the electrical length
divided by the square root of the permittivity; the latter is set to
its default value. The response is 0.29990704322;1.00062.
CORR:EDEL2?
Query the value of the delay at port 2. The delay is equal to the
electrical length divided by the speed of light in the vacuum, so
the response is 1.0006922856E-009.
CORR:LOSS2 2; LOSS2:FREQ 1.5 GHz; OFFS 3 dB
Define the offset loss parameters at port 2.
Manual operation:
See "Delay / Electrical Length / Mech. Length" on page 662
[SENSe<Ch>:]CORRection:EDELay<PhyPt>[:TIME] <Delay>
Defines the offset parameter for test port <PhyPt> as a delay time.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
Parameters:
<Delay> Delay
Range: -3.40282346638529E+038 s to
+3.40282346638529E+038 s.
*RST: 0 s
Default unit: s
Example: See [SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation: See "Delay / Electrical Length / Mech. Length" on page 662
SCPI command reference