Command reference
R&S
®
ZNA
1397User Manual 1178.6462.02 ─ 20
[SENSe:]SWEep:COUNt:ALL <Sweep>
Defines the number of sweeps to be measured in single sweep mode
(INITiate<Ch>:CONTinuous OFF). The setting is applied to all channels. Use
[SENSe<Ch>:]SWEep:COUNt to define the sweep count for a single channel.
Setting parameters:
<Sweep> Number of consecutive sweeps to be measured.
Range: 1 to 100000
*RST: 1
Example: See Chapter 8.2.4, "Data handling", on page 1611.
Usage: Setting only
Manual operation: See "Sweeps/Memory Size" on page 512
[SENSe<Ch>:]SWEep:AXIS:FREQuency <Scale>
Selects either the channel base frequency or one of the port frequencies as the stimu-
lus axis in all diagrams of the active recall set. This command has no effect if no port-
specific frequencies are defined.
Suffix:
<Ch>
.
Channel number
Parameters:
<Scale> String parameter, one of the following signals:
'Channel Base; Source' - base channel frequency
'Port 1; Source' - source frequency at port 1
'Port 2; Source' - source frequency at port 2
...
'Gen 1; Source' - external generator frequency 1
...
'Pmtr 1; Receiver' - external power meter frequency 1
...
*RST: 'Channel Base: Source'
SCPI command reference