Command reference
R&S
®
ZNA
1510User Manual 1178.6462.02 ─ 20
Example:
SYSTem:IDENtify:STRing 'MyDevice'; *IDN?
Define an identity string. The response is MyDevice.
SYSTem:IDENtify:FACTory; *IDN?
Re-activate the factory setting. The response is
Rohde-Schwarz,<instrument type>-<ports>,<stock
no.><serial no.>,<firmware version>.
Usage: Event
Manual operation: See "Define *IDN + *OPT..." on page 815
SYSTem:KLOCk <Boolean>
Locks or unlocks the local controls of the analyzer. This includes the front panel keys,
the keyboard, or other local interfaces.
Parameters:
<Boolean> ON | OFF - lock or unlock the local keys.
*RST: OFF
Example:
SYST:KLOCk ON
Lock the local keys.
SYSTem:LANGuage <Language>
Specifies the remote language for the analyzer.
Parameters:
<Language> Command syntax for the analyzer, string variable:
'SCPI' R&S ZNA-specific command set: the analyzer supports
all commands described in this documentation.
'ZVR' | 'ZVABT' compatibility with network analyzers of the
R&S ZVR and R&S ZVA/B/T families.
'PNA' | 'HP8510' | 'HP8720' | 'HP8753' | 'HP8714' | 'HP8530' |
'ENA' compatibility with network analyzers from other manu-
facturers.
*RST: n/a - a reset does not affect the language setting.
The factory setting is SCPI.
Example:
SYST:LANG 'PNA'
Select a PNA-compatible command set.
Manual operation: See "Remote Language" on page 814
SYSTem:LOGGing:REMote[:STATe] <Boolean>
Enables logging of all remote control commands transferred to the analyzer.