Command reference
R&S
®
ZNA
1524User Manual 1178.6462.02 ─ 20
TRIGger<Ch>[:SEQuence]:LINK <MeasSequence>
Selects the triggered measurement sequence. The identifier for the sequence is a
string variable.
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasSequence> Triggered measurement sequence, string variable.
'SWEep' – trigger event starts an entire sweep.
'SEGMent' – trigger event starts a sweep segment, if segmented
frequency sweep is active (see example below). If another
sweep type is active, the trigger event starts an entire sweep.
'POINt' – trigger event starts measurement at the next sweep
point.
'PPOint' – trigger event starts the next partial measurement at
the current or at the next sweep point.
*RST: 'SWEep'
Example:
SEGM:ADD; :SWE:TYPE SEGM
Select segmented frequency sweep.
TRIG:LINK 'SEGM'
Select a trigger segment as triggered measurement sequence.
TRIG:LINK?
Query the triggered measurement sequence. The response is
'SEGMENT'.
Manual operation: See "Sequence" on page 506
TRIGger[:SEQuence]:LTRigger[:STATe] <Boolean>
Activates/deactivates tracking of lost trigger events.
Parameters:
<Boolean> ON (1)
Lost trigger traclking active
OFF (0)
Lost trigger tracking inactive
Manual operation: See "Activate Lost Trigger" on page 809
TRIGger[:SEQuence]:LTRigger:COUNt?
Returns the number of lost trigger events the analyzer firmware observed since the last
reset of the lost trigger event log (see TRIGger[:SEQuence]:LTRigger:RESet
on page 1525).
Usage:
Query only
Manual operation: See "Sequence" on page 506
SCPI command reference