Command reference
R&S
®
ZNA
1524User Manual 1178.6462.02 ─ 20
TRIGger<Ch>[:SEQuence]:LINK <MeasSequence>
Selects the triggered measurement sequence. The identifier for the sequence is a
string variable.
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasSequence> Triggered measurement sequence, string variable.
'SWEep' trigger event starts an entire sweep.
'SEGMent' trigger event starts a sweep segment, if segmented
frequency sweep is active (see example below). If another
sweep type is active, the trigger event starts an entire sweep.
'POINt' trigger event starts measurement at the next sweep
point.
'PPOint' trigger event starts the next partial measurement at
the current or at the next sweep point.
*RST: 'SWEep'
Example:
SEGM:ADD; :SWE:TYPE SEGM
Select segmented frequency sweep.
TRIG:LINK 'SEGM'
Select a trigger segment as triggered measurement sequence.
TRIG:LINK?
Query the triggered measurement sequence. The response is
'SEGMENT'.
Manual operation: See "Sequence" on page 506
TRIGger[:SEQuence]:LTRigger[:STATe] <Boolean>
Activates/deactivates tracking of lost trigger events.
Parameters:
<Boolean> ON (1)
Lost trigger traclking active
OFF (0)
Lost trigger tracking inactive
Manual operation: See "Activate Lost Trigger" on page 809
TRIGger[:SEQuence]:LTRigger:COUNt?
Returns the number of lost trigger events the analyzer firmware observed since the last
reset of the lost trigger event log (see TRIGger[:SEQuence]:LTRigger:RESet
on page 1525).
Usage:
Query only
Manual operation: See "Sequence" on page 506