Command reference
R&S
®
ZNA
1525User Manual 1178.6462.02 ─ 20
TRIGger[:SEQuence]:LTRigger:RESet
Clears the lost trigger event log.
Usage:  Event
Manual operation:  See "Clear Lost Trigger Log" on page 809
TRIGger<Ch>[:SEQuence]:MULTiple:HOLDoff <MeasSequence>[, <DelayTime>]
Sets a delay time between the trigger event and the start of the measurement ("Trigger
Delay") in multiple trigger mode.
Suffix: 
<Ch>
.
Channel number
Parameters:
<MeasSequence> SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence, PPOint denotes "partial
measurement"; see TRIGger<Ch>[:SEQuence]:LINK.
<DelayTime> Delay time
Range:  0 s to 13680 s
Increment:  10 ms
*RST:  0 s
Default unit: s
Example:  See TRIGger<Ch>[:SEQuence]:MULTiple:SOURce<inp>
Manual operation:  See "... /Logic/ ..." on page 509
TRIGger<Ch>[:SEQuence]:MULTiple:LOGic <Sequence>[, <TrigLogic>]
Defines the logical combination of the trigger events defined in TRIGger<Ch>[:
SEQuence]:MULTiple:SOURce<inp> that will trigger measurement sequence
<Sequence>.
Suffix: 
<Ch>
.
Channel number
Parameters:
<Sequence> SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence
PPOint denotes "partial measurement"; see TRIGger<Ch>[:
SEQuence]:LINK.
<TrigLogic> NONE | AND | OR
NONE
only trigger event 1 is considered
AND
logical AND of trigger events 1 and 2
SCPI command reference