GUI reference
R&S
®
ZNA
513User Manual 1178.6462.02 ─ 20
"All Channels on Hold": all channels are set to "Single" sweep mode.
Restart Sweep initiates a new measurement cycle.
These actions are only available for the DEFAULT Remote Language). For other
remote languages, the Continuous / Single setting applies to all channels and the "All
Channels Continuous"/"All Channels on Hold" buttons are greyed out;
Remote command:
INITiate:CONTinuous:ALL
Sweep Controller
Activates/deactivates the (resizable) "Sweep Info" dialog, which displays the current
sweep stage. The "Sweep Info" dialog is particularly useful for long duration sweeps
that are executed in single sweep mode: by observing the dialog, it is easy to realize
when the sweep is done.
The possible sweep stages and how they are displayed partly depend on the selected
sweep mode (see "Continuous / Single" on page 511):
Sweep stage Sweep controller display
in "Continuous" sweep
mode
in "Single" sweep mode
No ongoing sweep Idle Idle
Sweep is being prepared Preparing Preparing
Ongoing sweep
1)
Continuous Sweeping
if #Sweeps = 1
Sweeping m/n
if n = #Sweeps > 1
Measurement results are being calculated Continuous calculation Calculation
1)
The VNA is sweeping or waits for a trigger signal.
5.10.4.2 Restart Manager dialog
The "Restart Manager" dialog defines the Restart Sweep behavior in "Single" sweep
mode.
Access: Channel – [Sweep] > "Sweep Control" > "Restart Manager..."